Copy link
Transcutaneous Pacing
Last updated: 02/04/2026
Key Points
- Transcutaneous pacing is a lifesaving, temporary measure to address symptomatic bradyarrhythmias.
- If pacing will be required for a prolonged period, transitioning to a transvenous or more permanent pacing mode should be sought out.
- Successful capture must be confirmed by electrocardiography.
- The most common limiting factors include incorrect electrode placement, poor electrode-skin contact, and patient discomfort.
Introduction1
- Paul Zoll first applied clinically effective temporary cardiac pacing in 1952, using two hypodermic needles to the chest to apply a pulsating current.
- Technological developments since then have led to endocardial, epicardial, and gastroesophageal approaches, as well as the refinement of external pacing.
- All approaches involve providing rate support via electrodes driven by an external pulse generator, which can be easily discontinued once the underlying cause is addressed.
Indications1
Table 1. Indications for emergency and elective temporary pacing.
Abbreviations: BBB, bundle branch block, AVB, atrioventricular block, MI, myocardial infarction; TAVR, transcatheter aortic valve replacement
Figure 1. American College of Cardiology Foundation/American Heart Association Guidelines for the Management of Acute MI – Bradyarrhythmias & Heart Block, Indications for Transcutaneous Pacing (Class I)2.
Abbreviations: MI, myocardial infarction; AV, atrioventricular, LAFB, left anterior fascicular block; LPFB, left posterior fascicular block
Contraindications
- If the underlying pathology causing the bradyarrhythmia is not considered to be transient.
- Severe hypothermia is the cause of bradycardia.
- Skin issues (i.e., burns, open wounds) at sites of pad attachment.
- Adjust the pad placement position.
- Relatively contraindicated in prolonged asystole (>20 mins) due to poor resuscitative outcomes in these patients.
Advantages of Transcutaneous Pacing
- Widely available in most crash carts, along with defibrillator units
- Easy to perform and requires minimal training
- It can be initiated almost instantly without the setup and insertion time required by invasive techniques.
- Carries less risk and is more cost-effective
Figure 2. Zoll E-Series (2). Flickr. 2010. Accessed November 28, 2025. https://www.flickr.com/photos/intropin/4499143572?utm_source=chatgpt.com
Approaches to Temporary Pacing1-3
External (Transcutaneous) Pacing
- Modern transthoracic pacemakers function strictly in demand mode, with a maximum output around 150 milliamperes (mA).
- Skin at pacing patches should be thoroughly cleaned and allowed to dry.
- Consider trimming hair for better pad-skin contact.
- Most modern-day defibrillators have transcutaneous pacing capacity.
- Prophylactic pad placement with defibrillation function in cardiac surgery is a frequent practice given the high risk of perioperative tachyarrhythmias.
- Common indications include redo sternotomies, low ejection fraction, preexisting bradycardia or atrioventricular block, and dilated cardiomyopathies.
- Prophylactic pad placement with defibrillation function in cardiac surgery is a frequent practice given the high risk of perioperative tachyarrhythmias.
- Pad placement should be standardized to one pad placed below the right clavicle, just to the right of the upper sternal border, and the other with its center in the left midaxillary line, below the armpit4 (Figure 3).
- If this position is not possible, placing a pad on the left side of the chest, between midline and the nipple, and another posteriorly, just below the left scapula, is considered good practice.4
- In female patients, it is recommended to place the anterior pad to the left of the lower sternum, avoiding breast tissue whenever possible.4
Figure 3. Transcutaneous pacing pad placement. Source: Ramzy M. Rebel EM. 2022. CC BY NC ND
- Transcutaneous pacing should be considered only in emergencies or for rescue treatment until recovery or until a more reliable transvenous pacer is placed.
Settings
- Pacing units will allow for adjustable pacing rate (beats per minute (bpm)) and ventricular output (mA), intended to trigger ventricular depolarization.
- Unlike more permanent pacing devices, the sensing threshold of intrinsic ventricular electrical activity is not adjustable.
- Pacing threshold is the minimum amount of output required to capture a ventricular depolarization or until patient discomfort is considered intolerable.
- In conscious patients, output should start low and slowly increase until the threshold is reached.
- In emergencies, output is started high (to assure capture) and then dropped until the threshold is reached.
- The set output should be 5-10 mA above this pacing threshold.
- It might be increased in patients with pericardial effusions, emphysema, or positive pressure ventilation (due to increased chest cavity impedance).
Assessment of Successful Pacing3
- Successful electrical capture is confirmed by inspecting the ECG tracing on the monitor.
- Widened QRS (ventricular depolarization), followed by a distinct ST segment and broad T wave
- The set pacing rate should match the manual pulse check and monitor heart rate.
- Consider a 12-lead ECG to document appropriate capture
Figure 4. Electrocardiogram showing pacemaker activity. P marks the normal atrial activity. Due to the block, it does not reach the ventricles, but is registered by the pacemaker, which sends stimuli (S) that activate the ventricles. Source: Gjesdal K. pacemaker. In: Store medisinske leksikon. snl.no. 2025. Accessed November 28, 2025. Link CC BY SA 3.0
Complications3
- Failure to capture
- Capture should be evaluated periodically since the pacing threshold can change.
- Suboptimal electrode placement is the most common cause.
- Confirm correct chest placement of electrodes (anterior vs posterior).
- Pads should be placed away from bony structures.
- Poor skin-electrode contact from sweat, debris, and hair. It can also present as increased pacing thresholds.
- Increased chest impedance.
- Due to fluids (i.e., effusions) or intrathoracic air (i.e., pneumothorax, positive pressure ventilation, recent thoracic surgery)
- Myocardial ischemia or metabolic derangements can also cause increased pacing thresholds.
- Pain or discomfort is a limiting factor, particularly if higher outputs are required.
- Consider analgesia or anxiolytics
- Other
- Hiccups due to diaphragmatic stimulation
- Skin burns from prolonged transcutaneous pacing
- If atrial rhythm is preserved, AV dissociation can occur if pacer is on purely demand mode.
References
- Gammage MD. Temporary cardiac pacing. Heart. 2000;83(6):715-20. PubMed
- Ryan TJ, Antman EM, Brooks NH, et al. 1999 update: ACC/AHA Guidelines for the management of patients with acute myocardial infarction: Executive summary and recommendations: A report of the American College of Cardiology/American Heart Association task force on practice guidelines. Circulation. 1999;100(9):1016-1030. PubMed
- Doukky R, Bargout R, Kelly RF, Calvin JE. Using transcutaneous cardiac pacing to best advantage: How to ensure successful capture and avoid complications. J Crit Illn. 2003;18(5):219-25. PubMed
- Bray, J. E., Smyth, M. A., Perkins, G. D., et al. (2025). Basic Life Support: 2025 International Liaison Committee on Resuscitation consensus on science with treatment recommendations. Resuscitation, 215(Supplement 2), 110808. PubMed
Copyright Information

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.